Technical Data
Pipette Module
|
Target Volume (μL)
|
Systematic Error (%)
|
Random Error (%)
|
Single Channel 50μL
|
1*
|
5
|
5
|
10
|
1.5
|
0.8
|
50
|
1
|
0.4
|
Single Channel 250μL
|
5
|
3
|
2.5
|
50
|
1.4
|
0.6
|
200
|
1.5
|
0.4
|
Single Channel 1000μL
|
100
|
1.0
|
0.3
|
200
|
0.8
|
0.3
|
1000
|
0.8
|
0.3
|
* dispensed into liquid.
Supported Tips
Manufacturer |
Product Code |
Description |
Thermo Fisher |
901-261 |
20µl, Clear, Sterile |
902-021 |
50µl, Clear, Sterile, Filtered |
902-261 |
50µl, Clear, Sterile |
903-021 |
200µl, Clear, Sterile, Filtered |
903-261 |
200µl, Clear, Sterile |
904-021 |
1000µl, Clear, Sterile, Filtered |
904-261 |
1000µl, Clear, Sterile |
Corning Axygen |
TTF-20-C-HTR-S |
20µl, Clear, Sterile |
TT-50-C-HTR |
50µl, Clear, Sterile |
TTF-200-C-HTR-S |
175µl, Clear, Sterile, Filtered |
TT-200-C-HTR-S |
200µl, Clear, Sterile |
TTF-1000-C-HTR-S |
1000µl, Clear, Sterile, Filtered |
TT-1000-C-HTR-S |
1000µl, Clear, Sterile |
Starlabs |
R1181-5810 |
50µl, Clear, Filtered |
R1128-5700 |
50µl, Clear |
R1128-0810 |
200µl, Clear, Filtered |
R1181-0700 |
200µl, Clear |
R1128-2810 |
1000µl, Clear, Filtered |
R1181-2700 |
1000µl, Clear |
Note: The systematic and random error values can be achieved only under strictly controlled conditions during test as per ISO 23783 Annex D. The best compatibility is achieved when combining i.prep pipette modules with supported tips. The systematic error and random error verification results, in tests, have been achieved using configurations available in the latest and i.prep firmware release. Due to continuous product development the systemic and random error values are subject to change without notice.